Olicom 9600 Instrukcja Użytkownika Strona 34

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 59
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 33
9600 Compressor Installation, Operation, and Maintenance
P/N 8040540 3-11
C
HELIX TECHNOLOGY CORPORATION
-
TI C
RY
OG
EN
ICS
Multiple On-Board Cryopump Connections
CAUTION
Make sure the Compressor power is OFF before making any
connections to the rear panel.
Helium Line Connections
CAUTION
The use of several compressors on a single manifold feeding a common
supply header and a common return header requires special
precautions. Contact CTI-CRYOGENICS for a review of the intended
installation and for specific technical instructions. The use of a 9600
compressor on a manifold with other CTI-CRYOGENICS compressor
models requires a reduction of the helium charge pressure to 200 - 210
psig charge pressure to avoid helium safety valves from inadvertently
venting. Refer to Section 4 - Operation and Section 5 - Maintenance
for more information.
1. Connect the Supply and Return lines to the 9600 Compressor as
described in Connecting/Disconnecting Helium Flex Lines
within this section.
2. Connect the Gas Return Line to the customer supplied helium
manifold and then to the GAS RETURN connector on the
On-Board Cryopump.
3. Connect the Gas Supply Line to the customer supplied helium
manifold and then to the GAS SUPPLY connector on the On-Board
Cryopump.
Przeglądanie stron 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 58 59

Komentarze do niniejszej Instrukcji

Brak uwag